Model:MQ-DT(H)-50
Benchtop Temperature Humidity Test Chamber
Benchtop Temperature Humidity Test Chamber
product name |
Benchtop high and low temperature humidity test chamber |
||||
Product model |
MQ-DT(H)15F-2 |
MQ-DT(H)30F-2 |
MQ-DT(H)50F-2 |
||
Nominal volume |
Volume unit/L |
15 |
30 |
50 |
|
Inner size |
Width × depth × height (mm) |
300×200×250 |
300×300×350 |
350×350×400 |
|
External size |
Width × depth × height (mm) |
460×875×740 |
550×975×1050 |
600×1000×1200 |
|
Performance |
Temperature range |
A,-20℃~150℃ B,-40℃~150℃ C,-70℃~150℃ D,-90℃~150℃ |
|||
Humidity range |
A,10%~98%RH B, 20%~98%RH C, 30%~98%RH |
||||
Temperature fluctuation |
±0.5℃ |
||||
Humidity fluctuation |
≤±2%RH |
||||
Temperature uniformity |
≤2.0℃ |
||||
Temperature deviation |
≤±2.0℃ |
||||
Humidity deviation |
Humidity>75%RH:≤+2,-3%RH; Humidity<75%RH:≤±5%RH |
||||
Heating rate |
≥2.0℃/ Min |
||||
Cooling rate |
≥2.0℃/ Min |
||||
Linear temperature control |
0.5~1.0℃/Min |
||||
Refrigeration method |
Compressor refrigeration |
||||
cooling method |
Air-cooled |
||||
The main components |
Control System |
Siemens programmable controller + touch screen + independent development program |
|||
Heating system |
Fully independent system, high efficiency stainless steel finned tube heater |
||||
Humidifying system |
Imported electronic humidity sensor |
||||
Cooling System |
Import Danfoss Compressors |
||||
Circulatory system |
Temperature and low noise motor, stainless steel centrifugal impeller |
||||
Use materials |
Carton material |
High-quality carbon steel plate, phosphating electrostatic spray treatment |
|||
Inner box material |
International standard SUS304# stainless steel mirror treatment |
||||
Insulation material |
Hard flame retardant polyurethane foam |
||||
Sealing material |
High and low temperature environment-friendly silicone rubber seal |
||||
voltage |
220±10% VAC,L+N+G |
||||
Communication Interface |
RS232 communication interface, USB data interface, Ethernet |
||||
Standard configuration |
Sample holder 1 set, test lead hole 1 set |
||||
Optional accessories |
Recorder, Remote Control Software, Insulated Sample Holder, Split Manipulator, Dry Air Purge |
||||
safety devices |
Leakage, short circuit, motor overheating, heater overheating, humidification, water shortage, compressor overpressure, overload, multiple overtemperature protection |
||||
Environmental requirements |
Ambient temperature: 5~35°C, relative humidity: ≤85%RH, no strong vibration, electromagnetic radiation, no dust and corrosive substances around |
||||
Note: 1. temperature performance parameters are in the ambient temperature of 25 °C, studio no load conditions, in accordance with GB/T5170.2-2008 test; 2. according to the specific requirements of users customized non-standard high and low temperature test box; 3. This technical information is subject to change without notice. |
Standard Compliance
1. IEC60068-2-2 Environmental testing for electric and electronic products Test methods -
Tests B: Dry Heat
2.MIL-STD-810F-501.4 High temperature test
3.GB/T 11158-2008: high temperature test chamber technical conditions
4.GB/T 5170.2-2008: electrical and electronic products environmental test
equipment test methods