Model:MQ-WTH-08
Walk-In Temperature Humidity Test Chamber
Walk-In Temperature Humidity Test Chamber
Technical indicators
Product name |
Walk-in environmental laboratory |
||||
Nominal volume |
Unit/m³ |
8 |
22 |
48 |
120 |
Inner size |
Width × depth × height (mm) |
2000×2000×2000 |
3000×3000×2500 |
4000×4000×3000 |
5500×5500×4000 |
Laboratory function selection |
High temperature, low temperature, high and low temperature, high and low temperature heat, constant temperature and humidity, high and low temperature, light, rain, salt spray corrosion, dust |
||||
Performance Index |
Temperature range |
A,-40℃~80℃ B,-70℃~80℃ C,-90℃~80℃ D,-120℃~80℃ |
|||
Humidity range |
20%~98%RH |
||||
Temperature fluctuation |
±0.5℃ |
||||
Humidity fluctuation |
≤±2%RH |
||||
Temperature uniformity |
≤±2.0℃ |
||||
Temperature deviation |
≤±2.0℃ |
||||
Humidity deviation |
Humidity>75%RH:≤+2,-3%RH; Humidity<75%RH:≤±5%RH |
||||
Heating rate |
≥2.0°C/Min |
||||
Cooling rate |
≥1.0°C/Min |
||||
Refrigeration method |
Compressor refrigeration |
||||
Cooling method |
Air-cooled/water-cooled |
||||
Main components
|
Control System |
Siemens programmable controller + touch screen + independent development program |
|||
Heating system |
Fully independent system, high efficiency stainless steel finned tube heater |
||||
Humidifying system |
Imported electronic humidity sensor |
||||
Cooling System |
Importer-type scroll compressor |
||||
Circulatory system |
Temperature and low noise motor, stainless steel centrifugal impeller |
||||
Main materials |
Carton material |
High-quality carbon steel plate, phosphating electrostatic spray treatment |
|||
Inner box material |
International standard SUS304#stainless steel plate |
||||
Insulation material |
Hard flame retardant polyurethane foam |
||||
Sealing material |
High and low temperature environment-friendly silicone rubber seal |
||||
Voltage |
380±10% VAC, 50Hz; 3L+N+G |
||||
Communication Interface |
RS232 communication interface, USB data interface |
||||
Standard configuration |
1 sample holder, 1 test lead hole, 1 silicone plug |
||||
Optional accessories |
Recorder, Remote Control Software, Insulating Sample Holder, Split Manipulator, Dry Air Purge, Workbench, Specimen Cart |
||||
safety devices |
Leakage, short circuit, motor overheating, heater overheating, humidification, water shortage, compressor overpressure, overload, multiple overtemperature protection |
||||
Environmental requirements |
Ambient temperature: 5 ~ 35 °C, relative humidity: ≤ 85% RH, no strong vibration, electromagnetic radiation, dust and corrosive substances around; |
||||
Note: 1.temperature performance parameters are in the ambient temperature of 25 °C, studio no load conditions, in accordance with GB/T5170.2-2008 test; 2.can be customized according to the specific requirements of non-standard laboratories; 3. This technical information is subject to change without notice. |