Model:MQ-2TS-100
Thermal Shock Test Chamber
Thermal Shock Test Chamber
Product name |
Two box type thermal shock test chamber |
|||||
Product model |
MQ-2TS100 |
MQ-2TS180 |
MQ-2TS300 |
MQ-2TS1000 |
||
Nominal volume |
Unit/L |
100 |
180 |
300 |
1000 |
|
Inner size |
Width × depth × height (mm) |
500×500×400 |
600×600×500 |
700×700×600 |
1000×1000×1000 |
|
External size |
Width × depth × height (mm) |
1020×1200×1850 |
1200×1500×1900 |
1350×1600×1980 |
2450×1800×1980 |
|
Basket size |
Width × depth × height (mm) |
350×350×250 |
450×450×350 |
550×550×450 |
800×800×800 |
|
Operating mode |
A.vertical basket B. horizontal basket type |
|||||
Performance index |
temperature range |
A,-75℃~200℃ B,-90℃~200℃ C,-120℃~200℃ |
||||
Impact range |
A,-55℃~160℃ B,-75℃~160℃ C,-85℃~160℃ |
|||||
Temperature fluctuation |
±0.5℃ |
|||||
Temperature uniformity |
≤±2.0℃ |
|||||
Temperature deviation |
≤±2.0℃ |
|||||
heating rate |
RT~+200℃,≤40Min |
|||||
Cooling rate |
A,+25℃~-75℃,≤60Min;B,+25℃~-90℃,≤80Min;C,+25℃~-90℃,≤100Min; |
|||||
Temperature conversion time |
≤15s |
|||||
Temperature recovery time |
≤5Min |
|||||
Exposure conditions |
High temperature exposure 30Mins; Low temperature exposure 30Mins |
|||||
Refrigeration method |
Compressor refrigeration |
|||||
Cooling method |
Air-cooled/water-cooled |
Satandard Compliance :
1.IEC60068-2-14 Environmental testing for electric and electronic products-Part 2: Test Methods-Test N: Change of temperature
2. MIL-STD-810F-501.4 High temperature test
3. MIL-STD-810F-502.4 Low temperature test
4. MIL-STD-810F-503.4 Temperature Shock
5.GB/T 10592-2008: high and low temperature test chamber technical condition.
6.GJB 360B-2009:test methods for electronic and electrical components 107 requirements for temperature shock tests
7.GB/T 5170.2-2008:test method for environmental test equipment of electric and electronic products Temperature test equipment